Specifications
Calibration and Preventive Maintenance 99Chapter 6
Adjustment Procedure
for Agilent E6171A
Modules
1. Agilent E6171A modules require the simple calibration fixture shown
in Figure 6-1. Build the test fixture by adding leads to an Agilent
0699-4532 resistor as shown. The resistor is a 100Ω, 0.1% tolerance,
4 Watt, low temperature coefficient precision resistor. Adding the
extra leads allows for 4-Wire Ohms resistance measurements. Keep
all lead lengths as short as possible.
2. When instructed by the ‘autoadj.tpa’ testplan, remove the ribbon
cable from the J2 connector on the front of the Agilent E6171A
module. Connect the test fixture ribbon cable to the J2 connector on
the Agilent E6171A. Refer to Figure 6-2.
Figure 6-1. Agilent E6171A Calibration Fixture