Calibration Commands
Calibration commands let you:
K Enable and disable the calibration mode
K Change the calibration password
K Calibrate the current and voltage programming and measurement, and store new calibration constants
in nonvolatile memory.
NOTE: If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error.
CALibrate:CURRent
This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
Command Syntax
CALibrate:CURRent[:SOURce][:DC][:POSitive]
Parameters
None
Examples
CAL:CURR CAL:CURR:SOUR:DC:POS
Related Commands
CAL:CURR:NEG
CALibrate:CURRent:MEASure:LOWRange
Agilent 66311B only
This command initiates the calibration of the low-range current measurement circuit.
Command Syntax
CALibrate:CURRent:MEASure[:DC]:LOWRange
Parameters
None
Examples
CAL:CURR:MEAS
Related Commands
CAL:CURR
CALibrate:CURRent:MEASure:AC
Agilent 66311B only
This command initiates the calibration of the high bandwidth (ac) measurement circuit.
Command Syntax
CALibrate:CURRent:MEASure:AC
Parameters
None
Examples
CAL:CURR:MEAS:AC