RS-232 Troubleshooting
If you are having trouble communicating over the RS-232 interface, check the following:
The computer and the dc source must be configured for the same baud rate, parity, number of data
bits, and flow control options. Note that the dc source is configured for 1 start bit and 1 stop bit
(these values are fixed).
The correct interface cables or adapters must be used, as described under RS-232 Connector. Note
that even if the cable has the proper connectors for your system, the internal wiring may be incorrect.
The interface cable must be connected to the correct serial port on your computer (COM1, COM2,
etc.).
SCPI Conformance Information
The Agilent 66111A and 66311B conform to SCPI Version 1995.0.
SCPI Conformed Commands
ABOR OUTP:PROT:DEL STAT:QUES:ENAB
CAL:DATA OUT:PROT:STAT STAT:QUES:NTR
CAL:STAT [SOUR]:CURR[:LEV][:IMM][:AMPL] STAT:QUES:PTR
DISP[:WIND][:STAT] [SOUR]:CURR[:LEV]:TRIG[:AMPL] SYST:ERR?
DISP[:WIND]:TEXT[:DATA] [SOUR]:CURR:PROT:STAT SYST:LANG
INIT[:IMM]:SEQ | NAME [SOUR]:VOLT[:LEV][:IMM][:AMPL] SYST:VERS?
INIT:CONT:SEQ | NAME [SOUR]:VOLT[:LEV]:TRIG[:AMPL] TRIG[:SEQ1 | :TRAN][:IMM]
MEAS | FETC:ARR:CURR[:DC]? [SOUR]:VOLT:PROT TRIG[:SEQ1 | :TRAN]:SOUR
MEAS | FETC:ARR:VOLT[:DC]? SENS:CURR[:DC]:RANG[:UPP] TRIG:SEQ2 | ACQ[:IMM]
MEAS | FETC[:SCAL]:CURR[:DC]? SENS:FUNC TRIG:SEQ2 | ACQ:SOUR
MEAS | FETC[:SCAL]:CURR:HIGH? SENS:SWE:OFFS:POIN TRIG:SEQ:DEF
MEAS | FETC[:SCAL]:CURR:LOW? SENS:SWE:POIN *CLS
MEAS | FETC[:SCAL]:CURR:MAX? SENS:SWE:TINT *ESE*ESE?*ESR?
MEAS | FETC[:SCAL]:CURR:MIN? STAT:OPER[:EVEN]? *IDN?
MEAS | FETC[:SCAL]:VOLT[:DC]? STAT:OPER:COND? *OPC*OPC?*OPT?
MEAS | FETC[:SCAL]:VOLT:HIGH? STAT:OPER:ENAB *PSC*PSC?
MEAS | FETC[:SCAL]:VOLT:LOW? STAT:OPER:NTR *RCL*RST
MEAS | FETC[:SCAL]:VOLT:MAX? STAT:OPER:PTR *SAV*SRE*STB?
MEAS | FETC[:SCAL]:VOLT:MIN? STAT:PRES *TRG*TST?
OUTP[:STAT] STAT:QUES[:EVEN]? *WAI
OUTP:PROT:CLE STAT:QUES:COND?
Non-SCPI Commands
CAL:CURR[:SOUR][:DC][:POS] OUTP:DFI[:STAT]
CAL:CURR[:SOUR][:DC]:NEG OUTP:DFI:SOUR
CAL:MEAS[:DC]:LOWR OUTP:PON:STAT
CAL:MEAS:AC OUTP:RI:MODE
CAL:LEV OUTP:TYPE
CAL:PASS SENS:CURR:DET
CAL:SAVE [SOUR]:DIG:DATA[:VAL]
CAL:VOLT[:DC] [SOUR]:DIG:FUNC
CAL:VOLT:PROT TRIG:SEQ2 | ACQ:COUN:CURR | :VOLT
DISP[:WIND]:MODE TRIG:SEQ2 | ACQ:HYST:CURR | :VOLT
MEAS | FETC[:SCAL]:CURR:ACDC? TRIG:SEQ2 | ACQ:LEV:CURR | :VOLT
MEAS | FETC[:SCAL]:VOLT:ACDC? TRIG:SEQ2 | ACQ:SLOP:CURR | :VOLT