User`s guide
163
F
Compatibility Language
Introduction
The Agilent 66311B dc source covered by this manual is programmatically compatible with the
HP/Agilent 6632A, 6633A, and 6634A dc power supplies. This means that by using COMPatibility
language mode you can program the Agilent 66311B over the GPIB using COMPatibility commands.
To switch from SCPI commands to COMPatibility commands (and vice versa), use the SYST:LANG
command, as documented in chapter 8. The language setting is saved in non-volatile memory.
Table F-2 summarizes the COMPatibility commands. You may need to refer to the Series 6632A, 6633A,
and 6634A Operating Guide (p/n 5957-6360) for complete information on the COMPatibility commands.
The rest of this appendix discusses the COMPatibility language status system, and the COMPatibility
language error codes.
NOTE: For complete information on the Compatibility programming language, order the
6632A/6633A/6634A Operating manual, p/n 5957-6360.
Table F-1. COMPatibility Power-on Settings
Command Setting Command Setting
DC 1 (ON) POL 1 (normal)
DLY 8 ms (fast)
80 ms (normal)
PON last stored value
DSP 1 (ON) RELAY 1 (close)
ISET 0.04 A (66311B)
0.02 A (6632B)
0.008 A (6633B)
0.004 A (6634B)
RLYPON 1 (close)
OCP OFF SRQ 0
OUT 1 (ON) UNMASK 0
OVSET MAX VSET 0 V