E – Line Voltage Conversion
162
white/red/grey
grey
orange
orange
white/red/grey
grey
orange
white/red/grey
grey
orange
orange
white/red/grey
grey
orange
(spare)
orange
(spare)
orange
white/red
white/black
white/brown
white/brown
white/black
white/red
red
black
white/violet
white/yellow
white/violet
white/yellow
white/violet
white/yello
w
white/violet
white/yellow
Figure E-1, Power Transformer AC Input Connections
Install the Correct Line Fuse
♦ Unscrew the line fuse cap from the rear panel and install the correct fuse.
For 100/120 Vac operation: 2 A slow-blow; part number 2110-0303
For 220/230 Vac operation: 1 A slow-blow; part number 2110-0007
♦ Mark the voltage setting that the unit has been set to on the rear panel label.
Close the Unit
♦ Replace the outer cover.