User`s guide

161
E
Line Voltage Conversion
WARNING: Shock Hazard. Operating personnel must not remove instrument covers.
Component replacement and internal adjustments must be made only by qualified
service personnel.
Open the Unit
Turn off ac power and disconnect the power cord from the unit.
Loosen the two screws on the rear bezel and remove the bezel (use a #15 Torx drive).
Remove the two screws on the bottom of the unit (use a #15 Torx drive).
Pull the cover back to remove it from the unit.
Configure the Power Transformer
Locate the ac input wiring harness on the left side of the transformer
Use a needle nose pliers and connect the ac input wiring harness according to the information in the
following figure: