User`s guide

Verification and Calibration - B
143
Front Panel Calibration Procedure
These procedures assume you understand how to operate front panel keys (see chapter 5). Make sure the
sense switch is set to Remote and the sense terminals are directly jumpered to the output terminals.
Calibration procedures can only be performed with the dc source set to the SCPI programming language.
Check the Language Setting
Action Normal Result
1. Turn on the dc source, access the Address menu and scroll to the
LANG: command.
LANG: SCPI or
LANG:COMP
2. If the language is set to COMPatibility, change it to SCPI and press
Enter. If the language is already set to SCPI, just continue with the
calibration procedure.
LANG: SCPI
Enable Calibration Mode
Action Display
3.
Reset the unit by selecting Output, scrolling to *RST and pressing
Enter.
*RST
4.
Press Output On/Off to enable the output.
00.003V 0.0006A
5.
To begin calibration press Shift Cal, scroll to CAL ON and press
Enter.
CAL ON 0.0
6.
Enter the calibration password from Entry keypad and press Enter. lf
the password is correct the Cal annunciator will come on.
If CAL DENIED appears, then an internal switch has been set to
prevent the calibration from being changed. (See the Service Manual.)
lf the password is incorrect, an error occurs. If the active password is
lost, the calibration function can be recovered by setting an internal
switch that defeats password protection. (See the Service Manual.)
CAL DENIED
OUT OF RANGE
Voltage Programming and Measurement Calibration
Action Display
7. Connect the DMM (in voltage mode) directly to the dc source. Do
not connect the load resistor or current shunt.
8.
Press Shift Cal, scroll to CAL VOLT, and press Enter.
CAL:VOLT
9.
Press Shift Cal, scroll to CAL LEV, and press Enter to select the
first calibration point.
CAL:LEV P1
10.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and
enter the voltage value displayed on the DMM.
CAL:DATA 0.00
11.
Press Shift Cal, scroll to CAL LEV, use รค to scroll to P2 (the
second calibration point), and press Enter.
CAL:LEV P2
12.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and
enter the second voltage value displayed on the DMM.
CAL:DATA 0.00