User`s guide

Verification and Calibration - B
141
Table B-2. Verification Test Record for Agilent Model 66111A
Model Agilent 66111A Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
10 mV
_____mV +10 mV
Front Panel measurement
V
out
5 mV
_____mV V
out
+5 mV
High Voltage V
out
14.982 V _______V 15.018 V
Front Panel measurement
V
out
9.5 mV
_______V V
out
+9.5 mV
Current Programming and Measurement
Low Current I
out
1.33 mA
_____mA 1.33 mA
High Current I
out
2.9972 A _______A 3.0028 A
Front Panel measurement
I
out
15 mA
_______A I
out
+15 mA
Current Sink Measurement
Low Current Sink measurement
I
sink
22.5 µA
_____mA
I
sink
+22.5 µA
High Current Sink measurement
I
sink
13 mA
_____A I
sink
+13 mA
Table B-3. Verification Test Record for Agilent Model 66311B
Model Agilent 66311B Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
10 mV
_____mV +10 mV
Front Panel measurement
V
out
5 mV
_____mV V
out
+5 mV
High Voltage V
out
14.982 V _______V 15.018 V
Front Panel measurement
V
out
9.5 mV
_______V V
out
+9.5 mV
Current Programming and Measurement (High Range)
Low Current I
out
1.33 mA
_____mA 1.33 mA
High Current I
out
2.9972 A _______A 3.0028 A
Front Panel measurement
I
out
6.5 mA
_______A I
out
+6.5 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5 µA _____µAI
out
+2.5 µA
High Current measurement
I
out
22.5 µA
_____mA
I
out
+22.5 µA
Current Sink Measurement
Low Current Sink measurement
I
sink
22.5 µA
_____mA
I
sink
+22.5 µA
High Current Sink measurement
I
sink
5.1 mA
_____A I
sink
+5.1 mA