User`s guide
A - Specifications
134
Supplemental Characteristics
Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical
performance determined either by design or type testing.
Table A-2. Supplemental Characteristics
Parameter Agilent 66111A Agilent 66311B
Average Current Measurement
Resolution
Hight Range:
Low Range:
3.5 mA
NA
213 µA
0.6 µA
Readback Accuracy Temperature
Coefficient (change/C°)
Voltage: 0.01% +
Current (dc): 0.02% +
Current (acdc): 0.05% +
Current (low range): 0.01% +
300 µV
15 µA
NA
NA
300 µV
15 µA
120 µA
0.3 µA
Maximum Current Measurement 7 A
Output Programming Range Voltage:
Current:
OVP:
0 – 15.535 V
0 – 3.0712 A
22 V
Average Programming Resolution Voltage:
Current:
OVP:
3.75 mV
0.67 mA
100 mV
Programming Accuracy
Temperature Coefficient
(change/C°)
Voltage: 0.01% +
Current: 0.01% +
OVP: 0.015% +
0.5 mV
15 µA
4 mV
OVP Accuracy
2.4 % + 240 mV
Drift
2
Voltage: 0.01% +
Current: 0.01% +
1 mV
30 µA
Sink Current - 2 A @ 7.5 V
1
Output Voltage Rise/Fall Time
(for a 10 – 90% or 90 – 10% change
of the total excursion)
< 200 µs
Output Voltage Settling Time
(time to settle within 1 LSB or
0.025% times the rated voltage of the
final value)
2 ms
Dynamic Measurement Accuracy
Instantaneous Voltage:
Instantaneous Current:
NA 0.03% + 5 mV
0.6% + 2 mA
3
Dynamic Measurement System
Buffer Length:
Sampling Rate Range:
NA 4096 points
15.6µs−31200s
1
The sink current decreases linearly from 2.8A @ 0 V to 1.2 A @ 15 V. The sink current does not track the programmed current.
2
Following a 30 minute warm-up, the change in output over 8 hours, under constant ambient, load and line operating conditions.
3
For full scale current changes with a risetime of 20 µs, an additional 0.5% error exists in the first data point in the buffer after
the change. The error percentage increases proportionally with the decrease in risetime.