User`s guide
133
A
Specifications
Specifications
Table A-1 lists the specifications of the dc source. Unless otherwise noted, specifications are warranted
over the ambient temperature range of 0 to 55 °C. Specifications apply with typical cellular phone
capacitive loads from 0µF to 12,000µF. Sensing is at the rear terminals of the power supply after a 30-
minute warm-up period. Sense terminals are externally jumpered to their respective output terminals.
Table A-1. Performance Specifications
Parameter Agilent 66111A Agilent 66311B
Output Ratings
Voltage:
Current:
Peak Current:
0 – 15 V
0 – 3 A
1
5 A
2
0 – 15 V
0 – 3 A
1
5 A
2
Programming Accuracy
(@ 25°C ±5°C)
Voltage: 0.05% +
+Current: 0.05% +
10 mV
1.33 mA
3
10 mV
1.33 mA
3
DC Measurement Accuracy
(via GPIB or front panel meters with
respect to actual output @ 25°C
±5°C)
Voltage: 0.03% +
High Current range
+20 mA to +rated I: 0.2% +
−20 mA to − rated I: 0.2% +
Low Current range
−20 mA to +20 mA: 0.1% +
5 mV
4
9 mA
9 mA
NA
5 mV
4
0.5 mA
5
1.1 mA
2.5 µA
Ripple and Noise
(in the range of 20 Hz to 20 MHz
with outputs ungrounded or with
either terminal grounded)
Voltage (rms/p-p):
Current (rms):
1 mV/6 mV
6
2 mA
Load Regulation
(change in output voltage or current
for any load change within ratings)
Voltage:
Current:
2 mV
0.75 mA
Line Regulation
(change in output voltage or current
for any line change within ratings)
Voltage:
Current:
0.5 mV
0.75 mA
Transient Response Time
< 35 µs
(For the output voltage to recover to 20 mV below its final value. Applies
to the negative portion of the output voltage transient waveform following
a 0.1 to 1.5 A load change in the High Mode compensation range.)
1
From 45° to 55° average current derates linearly from 3 A to 2.5 A. All other specifications are unaffected.
2
Peak current for up to a 7 millisecond time period. Average current over 11.7 milliseconds cannot exceed 3 A.
3
0.05% +2.13mA when programming between zero and 0.03% of full scale current.
4
Specification applies for output voltages greater than 10mV.
5
Applies with current detector set to DC. In ACDC mode the offset is 4 times the specified value for +current and 2.6 mA for
negative current.
6