User`s guide

Language Dictionary - 8
107
OUTPut:RI:MODE
This command selects the mode of operation of the Remote Inhibit protection. The RI mode is stored in
non-volatile memory. The following modes can be selected:
LATChing
causes a TTL low signal on the INH input to disable the output. The only way to
clear the latch is by sending OUTPut:PROTection:CLEAR while the INH input is
false.
LIVE
allows the INH input to disable the output in a non-latching manner. In other words,
the output follows the state of the INH input. When INH is low true, the output is
disabled. When INH is high the output is not affected.
OFF
the INH input is disabled.
Command Syntax
OUTPut:RI:MODE <mode>
Parameters
LATChing | LIVE | OFF
Examples
OUTP:RI:MODE LIVE
Query Syntax
OUTPut:RI:MODE?
Returned Parameters
<CRD>
Related Commands
OUTP:PROT:CLE
OUTPut:TYPE
This command programs the output compensation circuit. This circuit compensates the output of the dc
source according to the input capacitance of the phone being tested. The circuit covers the following
capacitance ranges:
LOW mode: 0 µF to 12,000 µF
HIGH mode: 5 µF to 12,000 µF
The Agilent 66311B is shipped with the output compensation set LOW mode. This mode provides the
stable operation for all types of phones. The HIGH mode setting provides better transient response
performance for phones that have input capacitances greater than 5 µF. In this mode however, the
operation of the dc source will be unstable when connected to phones that have input capacitances less
than 5 µF. (Units with option J01 are shipped with the compensation set to HIGH. The *RST value is
also HIGH.)
NOTE: If you set the output compensation to HIGH mode and you want the unit to power up in
HIGH mode, you must first save the desired output settings in non-volatile memory
location 0 with the *SAV command. Then program the dc source to power up with the
settings stored in memory location 0 with the OUTP:PON:STAT RCL0 command.
Command Syntax
OUTPut:TYPE[:CAPacitance] <setting>
Parameters
HIGH | LOW
*RST Value
LOW
Examples
OUTP:COMP HIGH
Query Syntax
OUTPput:TYPE[:CAPacitance]?
Returned Parameters
0 | 1