SENSe:FUNCtion
This command configures the measurement sensor to measure either voltage or current when an acquire
trigger is used. The query returns the function setting, either VOLT or CURR.
Command Syntax
SENSe:FUNCtion <function>
Parameters
"VOLTage" | "CURRent"
Examples
SENS:FUNC "VOLT"
Query Syntax
SENSe:FUNCtion?
Returned Parameters
<SRD>
SENSe:PROTection:STATe
This command enables or disables the open sense lead detection.
Command Syntax
SENSe:PROTection:STATe <bool>
Parameters
0 | OFF | 1 | ON
*RST Value
1
Examples
SENS:PROT:STAT ON
Query Syntax
SENSe:PROTect:STATe?
Returned Parameters
<NR3>
SENSe:SWEep:OFFSet:POINts
This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger. Positive values represent the delay after the trigger
occurs but before the samples are acquired.
Command Syntax
SENSe:SWEep:OFFSet:POINts <NRf+>
Parameters
-4095 through 2,000,000,000
*RST Value
0
Examples
SENS:SWE:OFFS:POIN -2047
Query Syntax
SENSe:SWEep:OFFSet:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR
SENSe:SWEep:POINts
This command defines the number of points in a data sweep.
Command Syntax
SENSe:SWEep:POINts<NRf+>
Parameters
1 through 4096
*RST Value
2048
Examples
SENS:SWE:POIN 1024
Query Syntax
SENSe:SWEep:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR