The default Sampling Frequency should be adequate for most
types of chromatography. For Frequency data sampling, the
data rate is selected in Hz (samples per second). For Period
sampling, you can select the number of seconds (or
milliseconds) between data points. The default is Frequency
and this is how most chromatography applications are set up.
Make sure that the Run Time is long enough for your last
expected peak to elute. If you do not know how long it will
take to elute, set the run time to a high number, such as 100
minutes. (You can stop the run manually after the last peak
has eluted.) After your first run you can then adjust the run
time to a more appropriate number.
Click the Trigger tab and select the Trigger Type for the type
of remote start (if any) you have installed for the instrument
you will be using. (If no trigger is configured, this tab will not
appear.) The trigger for each instrument is set up during
configuration.