User`s guide

76 User’s Guide
field. It is not possible to use an existing file name,
unless the file exists in located in a directory whose
path contains the term "public". For example, if you
data files are saved in a directory entitled
"C:\Public\Data", the files saved in this directory can
be overwritten. The software automatically appends a
.dat file extension.
Number of runs
Enter the number of runs you wish to make. The runs
will automatically proceed without review until
completed, incrementing each file name as
designated. If the sequence of single runs is aborted,
and the user then repeats the single acquisition
without changing any parameter, the run number will
start with the next number as if the sequence not
been aborted. For example, setting 4 runs with
starting run number of 101, then abort during run
102. When restarting, the next run number will be
105. If the Sample ID is also incremented, it will
increment in parallel.
Print Method Report
When this box is checked, the method report (or
reports) will be printed at the end of the run.
Amount Values
In this section, you can enter values that affect how
the concentrations are calculated. If you are making a
single data acquisition prior to calibrating your
method, simply leave these values at the default level.
Sample Amount
The Sample Amount value is used as a divisor during
calculation of concentrations. It is intended to
compensate for differences between samples due to
weighing and when percentages of the total sample
are being calculated rather than the amount detected
in an injection.