Technical data

9 Using the Reset to Factory Adapter for AIC Rev A/B
The Agilent Instrument Controller (AIC) (Rev A and B only) is
shipped with a USB flash drive that, when connected to an AIC, will
reset the AIC settings to the factory defaults. To reset an AIC to
factory defaults,
1. Turn on AIC. The AIC produces two short beeps when
initialization is complete.
2. When power up is done, plug the Reset to Factory
adapter into any USB port on the AIC.
3. The reset to factory will start.
4. The AIC will start beeping.
5. Remove the Reset to Factory adapter. You may remove it
as soon as the beeping starts.
6. The AIC will restart automatically and then proceed with
the rest of the reset to factory.
Important Notes for the Reset to Factory Adapter
The Reset to Factory Adapter will cause a reset to factory
defaults on any AIC, not just the one it was delivered with.
It can be reused any number of times.
Plug the Reset to Factory Adapter in AFTER the AIC is
powered on. Don’t plug it in before turning on the AIC.
The Reset to Factory Adapter must be connected to start the
reset to factory, but it must not be connected for the restarts
of the AIC that are involved in the automatic reset to factory
procedure.
After a reset to factory, the IP address of the AIC will be
reset to 192.168.1.1.