Technical data

7 Installing AIC Revision B for Agilent OL
Use this procedure for the Agilent Instrument Controller Revision B.
Revision B of the AIC is silver in color and is marked with a
“Revision B” sticker on the rear.
Install the Agilent Instrument Controller Network
Appliance
1. Before installing or configuring an Agilent Instrument
Controller, make sure all OL ECM and ICM components are
installed.
2. Verify which revision of the Agilent Instrument Controller
you have. Installation of Revision B (silver in color) is
described in this section. Instructions on using Revision A
(black in color) is described later in this guide.
3. Create an “AIC User” account. (See Add a New User in the
online help for how to do this.) This should be a unique user
with full administrator rights (call this user “aicuser” for
example).
4. From the Administration tab, click Global Administration,
and then right click Agilent Instrument Controller. In the
General tab of the Properties dialog box, enter the AIC user
account name and password, along with the account and
domain for the user. This will be used by the AICs to
communicate with the OL server. Click OK to close the
dialog box.
5. Disconnect the computer from the network and change the
TCP/IP settings to use the static IP address 192.168.1.2.
Contact a network administrator if you need assistance
doing this.
6. Connect one end of the network crossover cable (Part
Number G4690-60000) to the computer and the other end to
the main network port of the Agilent Instrument Controller if
more than one network port is available. (See Figure 1.)
Installing AIC Rev. B 15