Technical data
6 AIC 5000 Specifications
Dimensions
Height 3.95 inches 10.3 cm
Width 13.3 inches 33.78 cm
Depth 15.1 inches 38.35 cm
Approximate Weight 21 lb 9.53 kg
Temperature Range
Operating 50° to 95° F 10° to 35° C
Nonoperating -22° to 140° F -30° to 60° C
Relative Humidity (noncondensing)
Operating 10–90% 10–90%
Nonoperating (38.7° C max wet bulb) 5–95% 5–95%
Maximum Altitude (unpressurized)
Operating 10,000 ft 3048 m
Nonoperating 30,000 ft 9144 m
Note
Operating temperature is derated 1.0° C per 300 m
(1000 ft) to 3000 m (10,000 ft) above sea level, no
direct sustained sunlight. Maximum rate of change
is 10° C/Hr. The upper limit may be limited by the
type and number of options installed.
Power Supply 115 V 230 V
Operating Voltage Range 90–132 VAC 180–264 VAC
Rated Voltage Range 100–127 VAC 200–240 VAC
Rated Line Frequency 50–60 Hz 50–60 Hz
Power Output 185 W 185 W
Rated Input Current (maximum) 5 A @ 100 VAC 2.5 A @ 200 VAC
Heat Dissipation
Maximum 971 BTU/hr 245 kg-cal/hr
Typical (idle) 256 BTU/hr 65 kg-cal/hr
Installing the AIC 5000 13