Technical data
this button. A dialog box appears asking you if you want to
close the window. Click Yes.
9. Immediately disconnect the crossover cable from the Agilent
Instrument Controller and connect the laboratory network.
The AIC gives a short two-tone beep when initialization is
complete. If initialization fails, the AIC gives one or more
single longer beeps.
10. To verify the initialization was completed, log into OL as a
user with administrator privileges. From the Administration
tab, click Global Administration followed by Agilent
Instrument Controller.
11. Verify that the AIC appears. If it does not appear, wait 5
minutes, click on another node (such as Account
Administration), and then click Agilent Instrument
Controller again. If the AIC still does not appear, there are
several possibilities, given below. Once you have resolved the
problem, go back to the beginning of this section and repeat
the steps to set up the AIC.
• The address for the OL server was entered
incorrectly. On some networks, it may be necessary
to enter a fully-qualified name (for example
http://aspen.agilent.com).
• The laboratory network cable was not connected
soon enough after clicking Initialize (see step 7-9,
above).
Install the SS420x
If your instrument provides an analog detector signal, you must use
the SS420x analog to digital converter box. If you are using a digital
instrument, you do not need an SS420x.
Connect the SS420x interface to the Agilent Instrument Controller
using the serial communications cable provided. Connect one end of
the cable to the serial port on the back of the SS420x. Connect the
other end of the cable to any serial port on the Agilent Instrument
Controller (see Figure 2.) Make sure each connector is tightened.
6 Installing the AIC 5000