If selected incorrectly, you may see line frequency noise on your
chromatogram. For best performance, select a base frequency
close to the maximum sampling frequency you will be using to
acquire data.
Sampling frequencies available in the Acquisition Setup portion
of the method will reflect the Base Frequency selected.
Range
For each channel used, select the button next to the signal range
for the detector connected to that channel.
Information
This area is for information only. It contains manufacturing
information about the interface box that may be required in
service situations. If the box is not recognized, the “Connection
Status” field will display “Not Connected”. To update this
information, click the Refresh button.
When the dialog is completed, click OK.
SS420x Advanced Port Configuration
The Advanced button enables you to set the baud rate for the
SS420x interface. Click the Auto Detect button to determine
the current baud rate setting.
SS420x Acquisition Configuration
Each instrument using the SS420x for acquisition must be
configured to use the interface. This is done when configuring
SS420x User’s Guide 17