User`s guide

B - Verification and Calibration
58
Table B-5. Verification Test Record for Agilent 6613C
Model Agilent 6613C Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
20 mV
_______V +20 mV
Front Panel measurement
V
out
6 mV
_____mV V
out
+6 mV
High Voltage V
out
49.955 V _______V 20.045 V
Front Panel measurement
V
out
21 mV
_____mV V
out
+21 mV
Current Programming and Measurement (High Range)
Low Current I
out
0.5 mA
_______A 0.5 mA
High Current I
out
0.999 A _______A 1.001 A
Front Panel measurement
I
out
2.2 mA
_____mA I
out
+2.2 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5
µ
A _____
µ
AI
out
+2.5
µ
A
High Current measurement
I
out
22.5
µ
A _____
µ
AI
out
+22.5
µ
A
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5
µ
A _____
µ
AI
sin
k
+22.5
µ
A
High Current Sink measurement
I
sin
k
2.8 mA
_____mA I
sin
k
+2.8 mA
Table B-6. Verification Test Record for Agilent 6614C
Model Agilent 6614C Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
50 mV
_______V +50 mV
Front Panel measurement
V
out
12 mV
_____mV V
out
+12 mV
High Voltage V
out
99.900 V _______V 100.100 V
Front Panel measurement
V
out
42 mV
_____mV V
out
+42 mV
Current Programming and Measurement (High Range)
Low Current I
out
0.25 mA
_______A 0.25 mA
High Current I
out
0.4995 A _______A 0.5005 A
Front Panel measurement
I
out
1.1 mA
_____mA I
out
+1.1 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5 µA _____µAI
out
+2.5 µA
High Current measurement
I
out
22.5 µA _____µAI
out
+22.5 µA
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5 µA _____µAI
sin
k
+22.5 µA
High Current Sink measurement
I
sin
k
1.7 mA
_____mA I
sin
k
+1.7 mA