Changing the Deuterium Lamp
Y
ou should replace the deuterium lamp
w
hen the baseline noise and/or drift
exceed your application limits – or if the
lamp does not light. Use the following
procedure:
1. If the detector has been in use, turn it off
and allow it to cool before proceeding.
2. Remove the front cover.
3. Disconnect the electrical connection,
and unscrew the lamp to remove it
from the DAD.
4. Replace with a new lamp, taking care
not to touch the glass surface with
your hands.
5. Tighten the screws and reconnect the
lamp to its electrical connection.
6. Replace the front cover.
7. Reset the lamp counter and turn on
the lamp. Allow it to warm up for 10 - 20
minutes. Note: The new 'C' type
detectors use lamps with a chip installed
to automatically count used lamp time.
Maintenance Minder
When removing the lamp … unscrew both screws and unplug the lamp.
When installing the lamp … be sure to place the lamp into its auto-
aligning housing and screw snuggly.