Specifications
www.agilent.com/chem
•
800 227 9770
24
PUMP MAINTENANCE
Purge Valve – Exchanging the PTFE Frit 1100 standard pump modules
Purge valves feature a gold seal held in
p
lace with a plastic cap. Small leaks can
occur here, so you will need to replace the
gold seal after the valves have been
removed more than once or twice.
When replacing the frit, make sure that
t
he cutout side faces up, which provides
more surface area on the side facing the
solvent flow.
1. Remove capillary and waste tube.
2. Use a 14 mm wrench to
unscrew the valve.
3. Remove the plastic cap and
the gold seal.
4. Use tweezers to remove the old frit
and install a new frit.
5. Replace the gold seal and
the plastic cap.
6. Install the valve.
Pump Maintenance Parts
Maintenance Minder
Realign the purge valve waste tube in the
correct orientation during installation.
Description Part No. Price ($)
1100 Pump Start Up Kit includes: 1 outlet cap, 2 PTFE frits, 4 piston
seals, 1 outlet gold seal, 2 inlet filters, 1 cartridge for active inlet valve.
G1311-68710 444
PTFE Frits 5PK 01018-22707 26
Gold seal, outlet 5001-3707 25
Outlet caps, 4/PK 5062-2485 22
Purge valve assembly, SS For 1100 and 1050 pumps G1311-60009 351