Technical data
AGILENT dc ELECTRONIC LOADS 37
A
B
C
D
G
dc
Electronic
Loads
For more information in the U.S.A. call
1-800-452-4844
N3302A N3303A N3304A N3305A N3306A
Programming Resolution
Constant Current Mode 0.05mA/0.5mA 0.02mA/0.2mA 0.1 mA/1 mA 0.1mA /1 mA 0.2 mA/2 mA
Constant Voltage Mode 0.1 mV/1 mV 0.4 mV/4 mV 0.1 mV/1 mV 0.25 mV/2.5 mV 0.1 mV/1 mV
Constant Resistance Mode 0.07/0.7/7/70 m Ω 0.82/8.2/82 m Ω
0.035/0.35/3.5/35 m Ω 0.085/0.85/8.5/85 m Ω 0.0175/0.175/1.75/17.5 m Ω
Readback Resolution Current 0.05 mA/0.5 mA 0.02 mA/0.2 mA 0.1 mA/1 mA 0.1 mA/1 mA 0.2 mA/2 mA
Voltage 0.1 mV/1 mV 0.4 mV/4 mV 0.1 mV/1 mV 0.25 mV/2.5 mV 0.1 mV/1 mV
Programmable Slew Rate Current 0.2 A/ms to 0.017 A/ms to 0.1 A/ms to 0.1 A/ms to 0.2 A/ms to
2.5 A/µs 0.83 A/µs 5 A/µs 5 A/µs 10 A/µs
Voltage 0.1 V/ms to 1 V/ms to 0.1 V/ms to 1 V/ms to 0.1 V/ms to
0.5 V/µs 2 V/µs 0.5 V/µs 1.25 V/µs 0.5 V/µs
Ripple and Noise
(20 Hz to 10 MHz) Current 2 mA rms 1 mA rms 4 mA rms 4 mA rms 6 mA rms
20 mA p-p 10 mA p-p 40 mA p-p 40 mA p-p 60 mA p-p
Voltage 5 mV rms 12 mV rms 6 mV rms 10 mV rms 8 mV rms
Note: Specifications subject to change.
SUPPLEMENTAL CHARACTERISTICS
Supplemental Characteristics Continued
Analog Programming Bandwidth:
10 kHz (-3db frequency)
Analog Programming Voltage:
Voltage: 0-10V
Current: 0-10V
Analog Monitor Ports:
Voltage: 0-10V
Current: 0-10V
Remote Sensing:
5 V dc between sense and load input
Digital Inputs
Vil=0.9V max at Iil=-1mA
Vih-3.15V min (pull-up resistor on input)
Digital Outputs
Vol=0.72V max at Iol=1mA
Voh=4.4V min at Ioh=-20µA
Net Weight:
N3300A: 11.8kg (26lb); N3301A: 7.8kg (17lb)
N3302A, N3303A or N3304A: 3.2kg (7lb);
N3305A or N3306A: 5.4kg (13lb)
Shipping Weight:
N3300A: 15.9kg (35lb); N3301A: 9.8kg (22lb)
N3302A, N3303A, or N3304A: 4.5kg (10lb)
N3305A or N3306A: 7.3kg (16lb)
Option Descriptions
Opt. 800: Rack-mount kit for two N3301A
Mainframes mounted side-by-side
(p/n 5061-9694 and 5063-9215).
Opt. 908: Rack-mount kit (p/n 5063-9212 for a
N3300A, and p/n 5063-9245 for a N3301A)
Opt. 909: Rack-mount handles for N3300A
(p/n 5062-3984)
Opt. 910: Extra manual set, including one each of
the operating manual, programming reference
manual, and service manual. The programming
manual is available with the mainframes, and
therefore not the modules.
Note: Options 908, 909, and 800 require either
the slide kit (p/n 1494-0059) or slide rails to
support the weight of the load mainframe.
Accessory
p/n 34398A RS-232 Cable
p/n 5061-9694 Lock link Kit