Technical data

Visit our web site
http://www.agilent.com/find/power
66000A
66001A
66101A
66102A
66103A
66104A
66105A
66106A
24 AGILENT SYSTEM dc POWER SUPPLIES
Modular Power System:1,200W per mainframe
High density: eight slots in 7 inches of rack space
Output sequencing
Low ripple and noise
High-accuracy read-back of voltage and current over GPIB
Standard Commands for Programmable Instruments (SCPI)
Optional keyboard and display unit
Serial link to connect two mainframes at one GPIB address
Optional isolation and polarity-reversal relays
Built-in self-test
VXI
plug&play
drivers
(at 0˚ to 55˚ C unless otherwise noted)
SPECIFICATIONS
66101A 66102A 66103A 66104A 66105A 66106A
Output ratings Output Voltage 0 to 8 V 0 to 20 V 0 to 35 V 0 to 60 V 0 to 120 V 0 to 200 V
(at 40°C) Output Current 0 to 16 A 0 to 7.5 A 0 to 4.5 A 0 to 2.5 A 0 to 1.25 A 0 to 0.75 A
Output Power 128 W 150 W 150 W 150 W 150 W 150 W
Programming
accuracy Voltage 0.03%+ 3 mV 8 mV 13 mV 27 mV 54 mV 90 mV
(at 25°C ±5°C) Current 0.03%+ 6 mA 3 mA 2 mA 1.2 mA 0.6 mA 0.4 mA
Readback accuracy
(via GPIB or keyboard Voltage 0.02%+ 2 mV 5 mV 8 mV 16 mV 32 mV 54 mV
display at 25°C ±5°C) Current 0.02%+ 6 mA 3 mA 2 mA 1 mA 0.6 mA 0.3 mA
Ripple and noise Constant voltage rms 2 mV 3 mV 5 mV 9 mV 18 mV 30 mV
(20 Hz to 20 MHz) peak to peak 5 mV 7 mV 10 mV 15 mV 25 mV 50 mV
Constant current rms 8 mA 4 mA 2 mA 1 mA 1 mA 1 mA
Line regulation Voltage 0.5 mV 0.5 mV 1 mV 2 mV 3 mV 5 mV
Current 0.75 mA 0.5 mA 0.3 mA 0.1 mA 50 µA 30 µA
Load regulation Voltage 1 mV 1 mV 1 mV 2 mV 4 mV 7 mV
Current 0.5 mA 0.2 mA 0.2 mA 0.1 mA 50 µA 30 µA
Transient response time
Less than 1 ms for the output voltage to recover within 100 mV of its previous level following any step change in load
current up to 10 percent of rated current
Supplemental Characteristics (Non-warranted characteristics determined by design that are useful in applying the product)
Average resolution Voltage 2.4 mV 5.9 mV 10.4 mV 18.0 mV 36.0 mV 60.0 mV
Current 4.6 mA 2.3 mA 1.4 mA 0.75 mA 0.39 mA 0.23 mA
Output voltage
programming (OVP) 50 mV 120 mV 200 mV 375 mV 750 mV 1.25 mV
OVP accuracy 250 mV 500 mV 800 mV 1 V 1.5 V 2.5 V
66000 Modular Power System
The Agilent 66000 modular power system simplifies test-system
assembly, cabling, programming, debugging and operation. It is
ideal for ATE and production test environments, where it can sup-
ply bias power and stimulus to subassemblies and final products.
The modular power system saves rack space, the 7-inch-high (4-
EIA units) mainframe can accommodate up to eight dc power
modules.
Key Features
GPIB-programmable voltage and current
Programmable over-voltage and over-current protection
Self-test initiated at power-up or from GPIB command
Electronic calibration over GPIB or from keyboard
Over-temperature protection
Discrete fault indicator/remote inhibit (DFI/RI)
Five nonvolatile store-recall states per output
User-definable power-on state
A
B
C
D
E
System
dc Power
Supplies
66000A (mainframe)
66001A (keyboard)