Technical data

E3646A
E3647A
E3648A
E3649A
System
dc Power
Supplies
For more information in the U.S.A. call
1-800-452-4844
Dual-Output, Dual-Range
Low noise/excellent line and load regulation
GPIB/RS-232 standard
SCPI (Standard Commands for Programmable Instrument)
compatible
Front and rear output terminals
Overvoltage protection
Remote sense at rear output
(at 0˚ to 55˚ C unless otherwise specified)
SPECIFICATIONS
E3646A, E3647A, E3648A, E3649A
Agilent E3646A – 49A Dual Output
The new E3640A-series programmable dc power supplies are
60W/100W dual output dc power supplies with standard GPIB
& RS-232 interface. For general-purpose use, these programma-
ble power supplies offer the performance of system power sup-
plies at a surprisingly affordable price.
Clean and Stable Output
All models deliver clean and reliable power, dependable regula-
tion (0.01%) and fast transient response (<50usec to within
15mV). With 0.01% load and line regulation, Agilent E364xA
power supplies keep output steady when power line and load
changes occur. These linear supplies specify both normal mode
voltage noise and common mode current noise to assure less
interference with your DUT.
Standard remote interface
Standard GPIB and RS-232 interfaces, SCPI programming and
plug&play drivers for Agilent VEE and NI LabView make it easy
to program and integrate into automated test systems. You can
monitor the output terminals for voltage and current, and a query
command lets you read the programmed voltage and current.
Versatile Power
Agilent E364xA power supplies give you the flexibility to select
from dual output ranges. Both front and rear output terminals
are standard, and output load is protected against over-voltage
protection function. Remote sense is available to eliminate the
errors due to voltage drops on the load leads. New front panel
binding posts offer flexibility to use safety test leads, banana
plugs or stripped wires. Store and recall key enables you to save
and recall up to 5 frequently used operating states. For dual out-
put models, two outputs are electrically isolated to each other
and allows voltage tracking.
Ordering Information
STANDARD: 115 Vac +/- 10%, 47 to 63 Hz
Opt 0E3 230 Vac +/- 10%, 47 to 63 Hz
Opt 0E9 100 Vac +/- 10%, 47 to 63 Hz
Opt 1CM Rackmount Kit (p/n 5063-9243)
Opt 910 Extra Manual set
A
CCESSORIES:
E3619A 18 AWG power supply test leads
Stability: Voltage <0.02% + 2 mV; Current <0.1% + 1 mA
Temperature coefficient: 0.01% + 3 mV, 0.02% + 3 mA change per
degree C over operating range 0-40 degree C after 30 minute warm-up
(0.02% + 5 mV for output 2)
Warranty: Three years
Size: 213 mm W x 133 mm H x 348 mm D (8.4 in. x 5.2 in. x 13.7 in.)
Weight: E3646A, E3647A: 7.4 kg (16.3 lbs)
E3648A, E3649A: 9.5 kg (20.9 lbs)
dc Floating Voltage: Output terminals can be floated up to
±240 Vdc from chassis ground.
Remote Sensing: Up to 1V can be dropped in each load
lead. The drop in the load leads subtracts from the
voltage available for the load.
Settling Time: Less than 90msec for the output voltage to
change from 1% to 99% or vice versa following the
receipt of VOLTage or APPLy command via direct
GPIB or RS-232 interface.
Product Regulation:
Designed to comply with UL3111-1; certi-
fied to CSA 22.2 No. 1010.1; conforms to IEC 61010-1;
complies with EMC directive 89/336/EEC(Group1, Class A)
OVP accuracy: 0.5% + 0.5 V, activation time:
3 V,
<1.5 ms, and <3 V, <10 ms
Isolation: ±240 Vdc
A
B
C
D
E
AGILENT SYSTEM dc POWER SUPPLIES
23
Multiple-Output: 60 W to 100 W
E3646A E3647A E3648A E3649A
Number of Outputs 2 2 2 2
dc outputs Voltage/Current 0 to 8 V/3 A 0 to 35 V/0.8 A 0 to 8 V/5 A 0 to 35 V/1.4 A
0 to 20 V/1.5 A 0 to 60 V/0.5 A 0 to 20 V/2.5 A 0 to 60 V/0.8 A
Power (max) 60 W 60 W 100 W 100 W
Load and line regulation Voltage/Current <0.01% + 3 mV <0.01% + 3 mV <0.01% + 3 mV <0.01% + 3 mV
<0.01% + 250 µA <0.01% + 250 µA <0.01% + 250 µA <0.01% + 250 µA
Ripple and noise Normal-Mode Voltage <500 µVrms/5 mVp-p <1 mVrms/8 mVp-p <500 µVrms/5 mVp-p <1 mVrms/8 mVp-p
(20 Hz to 20 MHz) Normal-Mode Current <4.0 mArms <4.0 mArms <4.0 mArms <4.0 mArms
Common-Mode Current <1.5 µArms <1.5 µArms <1.5 µArms <1.5 µArms
Programming accuracy Voltage <0.05% + 10 mV (<0.1% + 25 mA for output 2)
(at 25˚ C ± 5˚ C) Current <0.2% + 10 mA
Readback accuracy Voltage <0.05% + 5 mV (<0.1% + 25 mV for output 2)
(at 25˚ C ± 5˚ C) Current <0.15% + 5 mA (<0.15% + 10 mA for output 2)
Program resolution Voltage/Current 5 mV/1 mA 5 mV/1 mA 5 mV/1 mA 5 mV/ 1 mA
Readback resolution Voltage/Current 2 mV/1 mA 2 mV/1 mA 2 mV/1 mA 2 mV/ 1 mA
Meter resolution Voltage/Current 10 mV/1 mA 10 mV/1 mA 10 mV/1 mA 10 mV/ 1 mA
Transient response time <50 µs for output to recover within 15 mV following a change in output current from full load to half load or vice versa.