Specifications

92
Application Notes:
Sequential Shunt Regulation
(AN 1293)
5965-7329E
Modern Connectivity -
Using USB and LAN I/O Converters
(AN 1475-1)
5989-0123EN
Solar Array Simulators (Continued)
More detailed specifications at
www.agilent.com/find/E4350
Application Specific DC Power Supplies tailored solutions for specific needs
Non-volatile memory can store a
maximum of 3500 points. The tables
(I-V curves) are easily stored and
recalled with an IEEE-488.2 com-
mand. The table(s) stored in this
memory will be retained when the
power is turned off. Volatile memo-
ry greatly increases the flexibility
by saving up to 30,000 points.
Multiple tables are easily accessed
with IEEE-488.2 command. These
tables will be erased after power
is removed.
In Table Mode, current and voltage
offsets can be applied to the select-
ed table to simulate a change in the
operating conditions of the solar
array.
3. Fixed Mode: This is the default
mode when the unit is powered on.
The unit has the rectangular I-V
characteristics of a standard power
supply, when an output capacitor is
added in this mode.
Specifications
(at 0
˚
to 55
˚
C unless
otherwise specified)
E4350B- E4350B- E4350B-
J03 J04 J06
Special Order Special Order Special Order
Option Option Option
Number of outputs 111
GPIB Yes Yes Yes
Output ratings
(Simulator and Table Modes)
Max. Power 480 W 480 W 480 W
Voc. Max. 52 V 47 V 74 V
Isc. Max. 10 A 11 A 7 A
Output ratings (for mixed mode)
Max Power 480 W 480 W 480 W
V rated 0 - 48 V 0 - 43.5 V 0 - 68 V
I rated 0.10 A 0 - 10 A 0 - 7 A
Programming accuracy at 25
˚
C ±5
˚
C
Voltage (Fixed Mode) 0.075% + 8 mV 0.075% + 8 mV 0.075% + 11.5 mV
Current (Simulator and Fixed Mode) 0.2% + 27.5 mA 0.2% + 30.5 mA 0.2% + 17.5 mA
Ripple and noise
from 20 Hz to 20 MHz
Voltage rms 16 mV 16 mV 19 mV
Voltage p-p 125 mV 125 mV 150 mV
Current rms 5.5 mA 6.5 mA 4 mA
Supplemental Characteristics
for all model numbers
Load Switching Recovery Time: < 5 µs when
switched from short circuit to variable
load, to within 1.5 A of an operating point
on the I-V curve.
Remote Sensing: Up to 2 V+ (Voc-Vmp).
Add 3 mV to the voltage load regulation
specification for each 1 volt change in the
positive output lead due to load current
change.
Analog Programming of Output Current
Input Signal: 0 to -4.0 V
Input Impedance: 20 k Ohms nominal
Shunt Regulation: Switching frequency
up to 50 kHz
Series Regulation: Switching frequency
up to 50 kHz
OVP and OCP: Overvoltage and overcur-
rent protection triggers in 100 us
Capacitive Load: In fixed mode, the maxi-
mum load capacitance (without causing
instability) is 2000 uF. In simulator and
table mode, it is unconditionally stable at
all capacitive loads.
Inductive Load: The maximum load
inductance (without causing instability)
is 200 µH
Software Driver:
VXIPlug&Play
Regulatory Compliance: Listed to UL3101,
certified to CSA 22.2 No. 1010.1, complies
with EN 61010-1.
RFI Suppression: Complies with CISPR-11,
Group 1, Class A
Size: 425.5 mm W x 132.6 mm H x
497.8 mm D (16.75 in x 5.25 in x 19.6 in)
Weight: Net, 25 kg (54 lb); shipping,
28 kg (61 lb)
Warranty: One year