Specifications

The Agilent N6700 Low-Profile
Modular Power System (MPS) is
a switching regulated, multiple-
output programmable DC power
supply system with the perfor-
mance of a linear power supply.
The N6700 is a flexible modular
platform that allows you to mix
and match 16 different DC power
modules to create a 1- to 4-channel
DC power system to optimized
performance, power and price
to match test needs. Test system
engineers can invest in high-
performance outputs where speed
and accuracy are needed, or pur-
chase basic performance outputs
for simple DC power requirements.
Small Size
The Agilent N6700 MPS uses an
advanced switching power supply
design that fits within 1U of rack
space. It has side air vents (no
top or bottom air vents) so other
instruments can be mounted directly
above or below it. (Requires rack
mount kit)
Protection Features
Each N6700 module is protected
against over-voltage, over-current,
and over-temperature. A fault con-
dition in one module can be detected
within 10 microseconds by other
modules so that they can be quickly
shut down to avoid hazardous
conditions on your DUT.
High Performance DC Power Supplies speed and accuracy for test optimization
Specifications
(at 0
˚
to 55
˚
C, and
derated above 40
˚
C)
N6751A N6752A N6761A N6762A
M
ore detailed specifications at
www.agilent.com/find/N6700
73
Small size: up to 4 outputs in 1 U of rack space
16 DC power modules: basic, performance and precision models
Fast output programming with active downprogramming
Ultra fast command processing time
Output sequencing and advanced triggering system
Optional LIST mode, built-in digitizer and disconnect relays
N6700B, N6710B, N6731-36B, N6741B-46B,
N6751A-52A, N6761A-62A
Low-Profile Modular Power System
50-100 W GPIB, LAN, USB
Output Ratings
Voltage 50 V 50 V 50 V 50 V
Current 5 A 10 A 1.5 A 3 A
Power 50 W 100 W 50 W 100 W
Programming Accuracy (at 23°C ±5°C)
Voltage high range 0.06% + 0.006% + 0.016% + 0.016% +
19 mV 19 mV 6 mV 6 mV
Voltage low range (5.5 V) N/A N/A 0.016% + 0.016% +
1.5 mV 1.5 mV
Current high range 0.1% + 0.1% + 0.04% + 0.04% +
20 mA 20 mA 200 µA 200 µA
Current low range 100 mA, @ 0 - 7 V N/A N/A 0.04% + 0.04% +
15 µA 15 µA
100 mA, @ 0 - 50 V N/A N/A 0.04% + 0.04% +
55 µA 55 µA
Readback Accuracy (at 23°C ±5°C)
Voltage high range 0.05% + 0.05% + 0.016% + 0.016% +
20 mV 20 mV 6 mV 6 mV
Voltage low range 5.5 V N/A N/A 0.016% + 0.016% +
1.5 mV 1.5 mV
Current high range 0.1% + 0.1% + 0.04% + 0.04% +
4 mA 4 mA 160 µA 160 µA
Current low range 100 mA, @ 0 - 7 V
2
N/A N/A 0.03% + 0.03% +
15 µA 15 µA
100 mA, @ 0 - 50 V N/A N/A 0.03% + 0.03% +
55 µA 55 µA
Output Ripple and Noise (PARD) (from 20 Hz to 20 MHz)
CV peak-to-peak
1
6 mV 6 mV 6 mV 6 mV
CV rms 1 mV 1 mV 1 mV 1 mV
Load Regulation
Voltage 2 mV 2 mV 0.5 mV 0.5 mV
Current 2 mA 2 mA 65 µA 65 µA
Line Regulation
Voltage 1 mV 1 mV 0.5 mV 0.5 mV
Current 1 mA 1 mA 30 µA 30 µA
1
For typical values, refer to Supplemental Characteristics.
2
Applies when measuring 4006 data points (SENSe:SWEep:POINts = 4096).
3