Specifications

170
Agilent 53131A Universal Counter
The two-channel 53131A counter
offers a bandwidth of 225 MHz with
a resolution of 10 digits per second
when measuring frequency or period.
Time interval measurements may
also be made with a resolution of
500 ps. An optional third channel
can be added to any counter to
provide frequency measurements
up to 3 GHz, 5 GHz, or 12.4 GHz.
Agilent 53132A Universal Counter
For applications requiring higher
resolution, the 53132A offers the
same features and functions as the
53131A, with up to 12 digits per
second frequency resolution and
150 ps time interval resolution.
If more accuracy is required, a
choice of optional timebases may
be added to any of the counters.
In addition, the 53132A offers
advanced arming modes for time
interval measurements.
Agilent 53181A RF Counter
Optimized for RF applications, the
single-channel 10 digits per second.
53181A measures frequency, period
and peak voltage. A digit-blanking
function easily eliminates unneces-
sary digits when you want to read
measurements quickly. For higher
frequency measurements, choose
an optional second channel. A self-
guided menu makes this counter
exceptionally easy to use.
Connectivity
For computer-controlled systems
applications, each counter includes
a standard GPIB interface with full
SCPI-compatible programmability
and a data transfer rate of up to
200 fully formatted measurements
Frequency Counters 53131A, 53132A, and 53181A
(Continued)
More detailed specifications at
www.agilent.com/find/frequencycounters
Frequency Counters
Standard and optional high
stability timebases
Standard Medium Oven High Oven Ultra High Oven
(0° to 50°C) Option 001 Option 010 Option 012
Temperature stability < 5 x 10
-6
< 2 x 10
-7
< 2.5 x 10
-9
< 2.5 x 10
-9
(referenced to 25° C)
Aging Rate
(after 30 days)
Per Day: < 4 x 10
-8
< 5 x 10
-10
< 1 x 10
-10
Per Month: < 3 x 10
-7
< 2 x 10
-7
< 1.5 x 10
-8
< 3 x 10
-9
Per Year: < 2 x 10
-8
Turn-on stability < 2 x 10
-7
< 5 x 10
-9
< 5 x 10
-9
vs. time (30 minutes) (Referenced to (Referenced to (Referenced to
2 hours) 24 hours) 24 hours)
Calibration Manual Adjust Electronic Electronic Electronic
Note: that power to the time base is maintained when the counter is placed in standby via the front panel switch.
The internal fan will continue to operate when in standby to maintain long-term measurement reliability.
per second. Measurements can be
transferred via USB with an optional
82357A GPIB to USB converter. The
standard RS-232 talk-only interface
provides printer support or data
transfer to a computer through
a terminal-emulation program.
IntuiLink, free PC connectivity
software allows you to easily
transfer measurement data and
images into familiar PC applications
like Microsoft Office Excel and Word,
with little or no programming. Use
the following link for additional and
downloading instructions.
www.agilent.com/find/intuilink
Optional High Frequency Connector Coupling Power Damage
Frequency range range and level
Channels sensitivity
Option 015
1
100 MHz BNC AC -27 dBm to 5 Vrms
1.5 GHz Channel to 1.5 GHz +19 dBm
Optional 030 100 MHz BNC AC -27 dBm to 5 Vrms
3.0 GHz Channel to 3.0 GHz +19 dBm
(100 MHz to
2.7 GHz)
-21 dBm to
+13 dBm
(2.7 GHz to 3 GHz)
Optional 050 200 MHz to Type-N AC -23 dBm to 25 dBm
5.0 GHz Channel 5.0 GHz +13 dBm
Optional 124 200 MHz to Type-N AC -23 dBm to 25 dBm
12.4 GHz Channel 12.4 GHz +13 dBm
1
Option 015 is available only for the 53181A