Specifications

163
DC Current
• 8 ranges: 100 nA to 1 A
• Up to 1,350 readings/s (5
1
/
2
digits)
• Maximum sensitivity: 1 pA
• 14 ppm 24-hour accuracy
AC Current
• 5 ranges: 100 µA to 1 A
• 10 Hz to 100 kHz bandwidth
• Up to 50 readings/second
• 500 ppm 24-hour accuracy
Frequency and Period
• Voltage or current ranges
• Frequency: 1 Hz to 10 MHz
• Period: 100 ns to 1 second
• 0.01% accuracy
• AC or DC coupled
Throughput
Maximum Reading Rates
• 100,000 readings/s
at 4
1
/
2
digits (16 bits)
• 50,000 readings/s at 5
1
/
2
digits
• 6,000 readings/s at 6
1
/
2
digits
• 60 readings/s at 7
1
/
2
digits
• 6 readings/s at 8
1
/
2
digits
Measurement System Speed
• 100,000 readings/s over
GPIB or with internal memory
• 110 autoranges/s
• 340 function or range changes/s
• Postprocessed math from
internal memory
Math Functions
The 3458A performs the following
math functions on measurements:
null, scale, offset, rms filter, single pole
filter, thermistor linearization, db, dbm,
% error, pass/fail LIMIT TESTING, and
statistics. Two math functions may be
used at one time.
8
1
/
2
Digit Multimeter 3458A (Continued)
More detailed specifications at
www.agilent.com/find/multimeters
Digital Multimeters
Abbreviated Technical
Specifications
DC Voltage
Range Full Maximum 1-Year*
Transfer Accuracy
Input
Scale Resolution Accuracy
10 min., tref ±0.5°C
Impedence
ppm of reading + ppm of range
Noise Rejection (dB)
1
AC NMR
2
AC ECMR DC ECMR
100 mV 120.00000 10 nV 9(5) + 3 0.5 + 0.5 >10 G
1 V 1.20000000 10 nV 8(4) + 0.3 0.3 + 0.1 >10 G
10 V 12.0000000 100 nV 8(4) + 0.05 0.05 + 0.05 >10 G
100 V 120.000000 1 µV 10(6) + 0.3 0.5 + 0.1 10 M ±1%
1000 V 1050.00000 10 µV 10(6) + 0.1 1.5 + 0.05 10 M ±1%
One-year specifications for NPLC 100 within 24 hours and ±1ºC of last ACAL, Tcal ±5ºC, MATH NULL, fixed range.
Add 2 ppm of reading additional error for Agilent factory traceability of 10 V DC to US NIST. Traceability error is
the absolute error relative to National Standards associated with the source of last external calibration. Transfer
specifications for NPLC 100, following 4-hour warm-up. Full scale to 10% of full scale. Measurements on the
1000 V range are within 5% of the initial measurement value and following measurement settling. Tref is the
starting ambient temperature. Measurements are made on a fixed range using accepted metrology practices.
*High stability (Option 002) ppm of reading in parentheses.
NPLC <1 0 90 140
NPLC 1 60 150 140
NPLC 10 60 150 140
NPLC 100 60 160 140
NPLC = 1000 75 170 140
1
Applies for 1 k unbalance in the LO lead and ±0.1% of the line frequency currently set for LFREQ.
2
For line frequency ±1%, ACNMR is 40 dB for NPLC 1, or 55 dB for NPLC 100.
For line frequency ±5%, ACNMR is 30 dB for NPLC 100.
DC Maximum Input
Rated Input Nondestructive
HI to LO ±1000 V pk ±1200 V pk
LO to guard ±200 V pk ±300 V pk
Guard to earth ±500 V pk ±1000 V pk