Specifications
159
General Specifications:
Power: 100/120/220/240V, ±10%
Power Line Frequency:
45 to 66 Hz, 360 to 440 Hz
Power Consumption: 25 VA peak
(10 W average)
Operating Environment: 0 to 55°C,
full accuracy to 80% RH, 40°C
Storage Environment: -40°C to 70°C
Size: 88.5 mm H x 212.6 mm W
x 348.3 mm D (4 in x 8.5 in x 14 in)
Weight: Net 3.6 kg (8.0 lb);
shipping 5.9 kg (13 lb)
Safety: Designed to UL-1244,
IEC-348, CSA
Order Information:
34401A Multimeter
34401A-1CM Rackmount Kit
34401A-A6J ANSI Z540
Compliant Calibration
34161A Accessory Pouch
34171A Input Terminal Connector
(sold in pairs)
34172A Input Calibration Short
(sold in pairs)
34131A Hard Transit Case
Low-Cost 6
1
/
2
Digit Multimeter 34401A (Continued)
More detailed specifications at
www.agilent.com/find/34401A
Digital Multimeters
100 mV 120.0000 100 µV 1 µV 10 µV 10 MΩ or >10 GΩ
1 V 1.200006 1 µV 10 µV 100 µV 10 MΩ or >10 GΩ
10 V 12.00000 10 µV 100 µV 1 mV 10 MΩ or >10 GΩ
100 V 120.0000 100 µV 1 mV 10 mV 10 MΩ
1000 V 1050.000 1 mV 10 mV 100 mV 10 MΩ
Input Protection: >1000 V on all ranges
100 mV 0.0030 + 0.0030 0.0040 + 0.0035 0.0050 + 0.0035
1 V 0.0020 + 0.0006 0.0030 + 0.0007 0.0040 + 0.0007
10 V 0.0015 + 0.0004 0.0020 + 0.0005 0.0035 + 0.0005
100 V 0.0020 + 0.0006 0.0035 + 0.0006 0.0045 + 0.0006
1000 V 0.0020 + 0.0006 0.0035 + 0.0010 0.0045 + 0.0010
60 Hz 6 300 1000
50 Hz 6 300 1000
DC Voltage Input Characteristics
Range Maximum Reading Resolution in Digits Input Resistance
(6
1
/
2
digits)
6
1
/
2
5
1
/
2
4
1
/
2
Low Cost 6
1
/
2
Digital Multimeter
– Abbreviated Technical Specifications
Measurement Accuracy: ±(% of reading + % of range)
Range 24 Hour 90 Day 1 Year
23°C ± 1°C 23°C ± 1°C 23°C ± 1°C
Maximum Reading Rate: (readings/s)
Power Line Resolution in Digits
Frequency
6
1
/
2
5
1
/
2
4
1
/
2
3 to 5 Hz 1.00 + 0.04 1.00 + 0.03
5 to 10 Hz 0.35 + 0.04 0.35 + 0.03
10 Hz to 20 kHz 0.06 + 0.04 0.06 + 0.03
20 to 50 kHz 0.12 + 0.04 0.12 + 0.05
50 to 100 kHz 0.60 + 0.08 0.60 + 0.08
100 to 300 kHz
*
4.00 + 0.50 4.00 + 0.50
*
Typically 30% of reading error at 1 MHz
AC Voltage (true rms) Measurement Accuracy: ±(% of reading + % of range); 1 year, 23°C + 5°C
Frequency Ranges Ranges
100 mV 1, 10, 100, 750 V