Specifications

Application Notes:
Agilent AN 372-1 Power Supply Testing
(AN 372-1)
5952-4190
Agilent AN 372-2 Battery Testing
(AN 372-2)
5952-4191
Increasing DC Power Supply Test System
Throughput with Agilent Technologies
N3300A DC Electronic Loads
5980-0233E
Agilent Zero Volt Electronic Load
5968-6360E
Making Fuel Cell AC Impedance
Measurements Utilizing
Agilent N3300A Series Electronic Loads
5988-5358EN
103
Multiple-Input: 150 W to 600 W (Continued)
N3300A N3301A
Operating temperature range 0°C to 55°C 0°C to 55°C
Input ratings
Operating range 100 - 250 Vac 100 - 250 Vac
48 - 63 Hz 48 - 63 Hz
Input Current 4.2 A @ 100 - 127 Vac 2.3 A @ 100 - 250 Vac
2.2 A @ 200 - 250 Vac
Input VA 440 VA 230 VA
Inrush Current 38 A 18 A @ 115 Vac
36 A @ 230 Vac
More detailed specifications at
www.agilent.com/find/N3300
DC Electronic Loads maximize throughput with real-life loading conditions
Table A-3
Supplemental Characteristics
Supplemental Characteristics
for all model numbers
Analog Programming Bandwidth:
10 kHz (-3 db frequency) in CC mode only
Analog Programming Voltage:
Voltage: 0 - 10 V
Current: 0 - 10 V
Analog Monitor Ports:
Voltage: 0 - 10 V
Current: 0 - 10 V
Remote Sensing:
5 V DC between sense and load input
Digital/Trigger Inputs
Vil = 0.9 V max at Iil = -1 mA
Vih - 3.15 V min
(pull-up resistor on input)
Digital/Trigger Outputs
Vol = 0.72 V max at Iol = 1 mA
Voh = 4.4 V min at Ioh = -20 µ A
Agilent Models: N3300A
425.5mm
16.75"
624.8mm
24.6"
To p
177.8mm
7.0"
50.8mm
2.0"
190.5mm
7.5"
12.7mm
0.5"
Side
Rear