Technical data

Methods with Acquisition Parameters 3
Saving a method with data acquisition parameters
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 95
Table 10 System scripts and the actions they enable
Script name Actions the script enable
SCP_AcquireCalibrantData Sets “Cal/Ref Mass” to “Cal B” and
“LCStream” to “LC->Waste”. Allows you to
acquire data for the calibrant solution itself.
To be used only as the Pre-run script for
a method. The script itself does not do a run
and only augments an existing method. Do
not use as a standalone script in a worklist
run.
SCP_MSDivertValveToMS Sets the solvent divert valve to the “MS”
position.
SCP_MSDivertValveToWaste Sets the solvent divert valve to the “Waste”
position.
SCP_InstrumentStandby Puts the instrument in standby mode. This is
the same as clicking the “Standby” button on
the Instrument Status window.
SCP_LoadIdleMethod
(MethodName)
Loads a method to put the system in an idle
state. The first parameter is the method
name.
SCP_MSRefOff Turns off the MS reference ion solution (Ref
A)
SCP_PumpsAllandMSRefOff Turns off LC pumps and turns off the MS
reference ion solution.
SCP_PumpsAllOff Turns off all pumps.
SCP_LCCondition
(MethodName, LCStream)
Starts a run for conditioning the LC part of
the instrument. No data acquisition happens.
SCP_CalibrateTOFMassAxis Starts a TOF/Q-TOF mass axis calibration in
the active polarity using the default mass list
SCP_TraceOnOff Switches the trace to ON or OFF. The first
parameter must be either “on” or “off”.
SCP_ClearTrace Clears the trace file.