Technical data
84 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
3 Methods with Acquisition Parameters
TOF and Q-TOF acquisition parameters
The software sorts a list of possible precursor ions whose order
depends on the boundary parameters entered:
a Passes all ions through (TOF only) and sorts the list from
highest abundance to lowest (or by charge then
abundance)
b Excludes those masses in the specified mass range
c Sorts the list based on the priority of charges
d Moves preferred ions to the top of list in the order
specified
e Chooses the top ions on the list based on the entry for
maximum number of precursor ions per cycle (Max
Precursor per cycle)
f Excludes masses after a specified number of spectra have
been acquired and releases the exclusion after a specified
time (Active Exclusion)