Technical data
82 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
3 Methods with Acquisition Parameters
TOF and Q-TOF acquisition parameters
Spectral Parameters These parameters are the same as those
for the TOF mode, but they are applied to both the quad (MS)
and the TOF components (MS/MS).
Collision Energy You can enter multiple fixed collision energies
(each collision energy is used), the slope and offset of a line, or
a table of collision energies.
For general method development you set the slope and offset for
a curve whose collision energy the system selects to match the
precursor ion, m/z, on the curve.
Figure 36 The Spectral Parameters tab for Targeted MS/MS mode
Figure 37 The Collision Energy tab for Targeted MS/MS mode