Technical data
74 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
3 Methods with Acquisition Parameters
Automatic TOF and Q-TOF parameter changes during a run
Per time segment set points are labeled with (Seg). Per
experiment set points are labeled with (Expt). All other set
points are per run. These set points can be found in these tabs.
• General Tab
• Source Tab
• Acquisition Tab
Acquisition tab
You can set up time segments for changing these parameters.
MS mode The MS mode spectral parameters can be set for
each time segment.
Auto MS/MS mode (Q-TOF only) The parameters on the
following tabs can be set for each time segment:
• Q-TOF Spectral Parameters Tab
• Q-TOF Collision Energy Tab
• Q-TOF Precursor Selection I Tab
• Q-TOF Precursor Selection II Tab
• Q-TOF Preferred/Exclude Tab
Targeted MS/MS mode (Q-TOF only) The parameters on the
following tabs can be set for each time segment:
• Q-TOF Spectral Parameters Tab
Figure 30 Location for setup of Time Segments and Experiments