Technical data
Methods with Acquisition Parameters 3
Automatic TOF and Q-TOF parameter changes during a run
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 73
Automatic TOF and Q-TOF parameter changes during a run
You can set some TOF and Q-TOF method set points to be
different at different points in time during the run. The
different time points in the method are called time segments.
Within each time segment, you can have different sets of
parameters, called experiments.
Many parameters can be changed for different time segments.
Some parameters can be changed for different experiments in
the same time segment. For example, the gas temperature,
drying gas flow rate, and the nebulizer pressure must be the
same for all experiments in the same time segment, but
fragmentor and capillary voltages can be different for different
experiments.