Technical data
Instrument Preparation 2
Real-time parameter values (Actuals)
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 65
WPS, µWPS, h-ALS,
h-ALS-SL and
h-ALS-SL+
The same as ALS except no Vial and addition of Drawn
Volume, Sample Position and Needle Position
Binary Pump Solvent Ratio A, Solvent Ratio B, Flow, Pressure, Ripple,
Fill A,B, A1, B1, A2, B2, Power, Channel Name A, B,
Solvent Selection A, B
Capillary Pump Same as binary pump with no Fill A1, A2, B1, B2 and
with the addition of Solvent Ratio C, Solvent Ratio D, Fill
C, Fill D, Primary Flow, EMPV, Purge Status, Purge
Channel, Purge Time and Pump Op Mode
Nanopump Same as capillary pump with addition of FSAC State and
FSAC Step. FSAC stands for Flow Sensor Accuracy
Calibration. When you click the Accuracy Calibration
shortcut menu, these values show you the status of the
step in the test procedure.
Thermostatted
Column Compartment
Left Temp, Left Temp Set, Right Temp, Right Temp set,
Column Valve, Therm Power
35900E Current Vial
DAD G1315C,
G1315D, G4212A,
G4212B, G1365C and
G1365D
Sample Wl A-H, Sample Bw A-H, Reference On A-H,
Reference Wl A-H, Reference Bw A-H, UV Lamp, Vis
Lamp
DAD G1315A,
G1315B, G1365A, and
G1365B
Sample Wl A-E, Sample Bw A-E, Reference On A-E,
Reference Wl A-E, Reference Bw A-E, UV Lamp, Vis
Lamp
TOF or Q-TOF Rough Vacuum, High Vacuum, Gas Temp, Vaporizer
Temp, Drying Gas Flow, Nebulizer Pressure, Capillary
Current, Chamber Current, Corona Voltage, Charging
Voltage, Control State, Cal/Ref Mass, LC Stream
Table 4 Actuals available for display
Module Parameter or State