Technical data
Instrument Preparation 2
Storage and retrieval of tune results and Instrument Mode
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 61
Extended Dynamic Range (2 GHz) In this mode, the system
acquires data at a 4 GHz ADC rate in dual channel mode. One
channel is recorded at a high detector gain while the other
channel operates at a low detector gain. The firmware stitches
the two channels together to produce a scan with a sampling
rate of 2 GHz, which results in a greatly increased dynamic
range. The Mass Range can only be set to Low (1700 m/z) or
Standard (3200 m/z) if the Instrument Mode is set to this value.
Extended Mass Range (2 GHz) In this mode, the system acquires
data at a 2 GHz ADC rate. The Mass Range can be set to Low
(1700 m/z), Standard (3200 m/z) or High (10,000). This is the
only mode that you can use with the highest mass range.