Technical data

60 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
2 Instrument Preparation
Storage and retrieval of tune results and Instrument Mode
Storage and retrieval of tune results and Instrument Mode
You can store the tuning parameters in a single file (*.tun) using
the Save or Save As buttons in the Instrument State tab. You
can also load tune files in the Instrument State tab (Figure 23).
The Mass Range, the Fast Polarity Switching option, the Slicer
Mode, and the Instrument Mode are also stored in the tune file.
Instrument Modes
You select from four different instrument modes.
High Resolution (4 GHz, High Res Mode) In this mode, the
system acquires data at a 4 GHz ADC rate while special
processing is performed in real time on the mass peaks detected
in each transient, weighting the apex data of the mass peaks
much more heavily than the shoulders. This leads to narrower
peaks in the summed transients (the mass spectra) and hence
improves resolution. The Mass Range can only be set to Low
(1700 m/z) or Standard (3200 m/z) if the Instrument Mode is set
to this value.
4 GHz (High Resolution Mode disabled) In this mode, no
processing is used at the transient level; however, the system
acquires data at a 4 GHz ADC rate. The Mass Range can only be
set to Low (1700 m/z) or Standard (3200 m/z) if the Instrument
Mode is set to this value.
Figure 23 Instrument State tab