Technical data

48 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
2 Instrument Preparation
Column equilibration and conditioning
Column equilibration and conditioning
You can set up to equilibrate or condition a column in different
ways with the Data Acquisition program.
Equilibration Column equilibration eliminates any previously separated
compounds or impurities from the column after runs with
solvent of a single composition. To equilibrate a column before a
sample run, you pass the solvent that you intend to use for the
run through the column for a period of time.
Conditioning Column conditioning returns column characteristics to their
initial state after a gradient run. To condition a column before a
sample run, you pass the solvent of initial composition through
the column for a period of time.
Equilibration
You can equilibrate a column in one of three ways:
Thermostat. Column
Compartment (TCC)
Turn the right or left controller on or off Right-click the device panel and click
Configuration
Set the maximum controller temperatures Right-click the device panel and click Set
Temperature
Enter information for the instrument columns Tools > Analytical Column Setup
Diode Array Det. (DAD) Turn UV or Visible lamp on Right-click the device panel and click UV
Lamp On; Vis Lamp On
Set the date or time for automatic DAD turnon Right-click the device panel and click Control
Calibrate the DAD wavelength Right-click the device panel and click
Calibration
Bring the baseline of the plot to zero Right-click the device panel and click
Balance
Show an intensity plot for the detector Right-click the device panel and click
Intensity plot
Table 1 Tasks to set up the LC modules (continued)
If you have this module: And you want to: How: