Technical data

Instrument Preparation 2
LC module setup
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 47
Table 1 Tasks to set up the LC modules
If you have this module: And you want to: How:
Autosampler Change volumes for the installed syringe Done in the Device Configuration user
interface
Well-Plate Sampler
(WPS) or µWPS or
h-ALS or h-ALS-SL or
h-ALS-SL+
Change volumes for the installed syringe Done in the Device Configuration user
interface
Select tray type and its position Right-click the device panel and click Assign
Wellplates, Edit Wellplate Types
Reset injector, Move Home, Needle Up or Down,
set the valve mainpass or the bypass
Right-click the device panel and click Reset
Sampler, Home Arm, Needle Up/Down,
Switch Valve to Bypass/Mainpass, Switch
Thermostat On, Wash Needle
HTC/HTS Update the plate assignments on the computer Right-click the device panel and click Update
Plate Assignment
Reset the injector Right-click the device panel and click Reset
injector
Binary, isocratic, and
quaternary pumps
Turn the pump on, off or place in Standby Right-click the device panel and click Switch
Pump On/Off
Set the date and time for automatic pump turnon Right-click the device panel and click Control
Set up to monitor solvent levels Right-click the device panel and click Bottle
Fillings
CapPump Do same tasks as binary pump Same menu items as binary pump
Purge the pump Right-click the device panel and click Purge
On/Off
Set mixer and filter volumes Right-click the device panel and click
Configuration
Enable a fast change in solvent composition Right-click the device panel and click Start
Fast Composition Change
Do a flow sensor calibration Right-click the device panel and click
Calibration
Nanopump Do the same tasks as CapPump Same menu items as CapPump except for
Accuracy Calibration (flow sensor)