Technical data

Installation 1
Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide 9
Installation
Step 1. Prepare to install
1 Check that you have these parts:
Bundled LC/MSD Multimode ESI/APCI Source (p/n G1978A)
LC/MSD Multimode ESI/APCI Source (p/n G1978-65239)
LCMSD-TOF MM ESI/APCI Enablement Kit (p/n G1978-60251)
TOF Software A.02.02 or higher
TOF Multimode Installation Guide (p/n G1978-90007)
2 Check that the TOF instrument has serial numbers greater than
US50600501 or US54100700.
3 Check that the TOF has Smart Card 4.
4 Check that you have the supported revision of the TOF Software (and any
needed patch) or the MassHunter Workstation Software - Acquisition for
TOF software. Do not install the patch that comes with the ESI/APCI
Enablement Kit.
5 Check that you have these tools, supplies and chemicals. The items in this
list are not provided with your multimode source.
Cloths and gloves, clean, lint-free
Water and organics, such as acetone, methanol, acetonitrile or isopropyl
alcohol, all HPLC grade.
open-end wrench
Torx drive T10
6 Verify the current performance of the TOF system:
With the dual-ESI source installed, do Positive and Negative mode
Autotunes.
Examine the tune reports to verify that the TOF system meets tune
specifications.
7 In the MS TOF tab, click Parameters and select Positive ion polarity. Click
Save As to save the positive tune parameters as d:\Program Files\Agilent\TOF
Software\Tune\PositiveESI.prm.