Technical data

3 Installation Verification
38 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
Step 1. Auto tune
MassHunter TOF Acquisition for TOF A.02.02
Before the TOF system with multimode source can be used, the installation
and verification tasks must be completed. Auto tune is done in mixed mode,
with the TOF G1969-85000 ESI-L Low Concnetration Tuning Mix.
1 Click File > Open > Method.
2 Load the MMIautotune.m file.
3 Click the Tune tab and select Autotune.
The instrument will generate a positive Autotune report for the multimode
source. See Figure 28 on page 39.