Technical data

2 TOF Software and TOF Firmware Upgrade for Multimode
36 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
Step 3. Pump down the LC/MSD TOF and condition the high
voltage electronics
Before you try tuning the LC/MSD TOF, the high voltage electronics need to be
conditioned.
1 Click Start > Programs > Agilent > TOF Software > Tools > TOF
Diagnostics.
2 Click Connection > Connect.
By default, the IP address 192.168.254.12 for the LC/MSD TOF appears.
3 Click OK.
Once you are connected, the MS Status line will say Instrument
Responding. Also, the connection box at the bottom right of the screen will
be green.
4 In the HV Conditioning group box, select the 0.6 Hour Cycle (Quick Vent )
option, since the TOF instrument was vented under nitrogen. Select the
Condition HV button to initiate the high voltage conditioning process.
Tuning and Calibration cannot proceed until this process has completed.