Step 2. Upgrade the TOF firmware
1 Install the TOF Firmware Update tool. Use File Explorer and open the
CD-ROM drive for the Installation CD.
2 Open the folder Support\TOF Firmware.
3 Double-click MSFirmwareUpdate_1.16.exe.
The WinZip self-extractor dialog appears. Use the default installation path.
4 Click Unzip.
5 Click OK when you are informed that the files have been unzipped.
6 Click Close to complete the installation.
7 Double-click tofupdate.exe in the copied directory.
8 Enter the IP address of the TOF. By default, this is 192.168.254.12
9 Enter Y when prompted to continue the update.
10 When the TOF firmware has been updated, press the reset button on the
communication card (SmartCard 4) to reboot it. The reset button can be
accessed via the small opening above the SmartCard 4 serial port.
11 After the SmartCard 4 has finished initializing, start the TOF Software.
12 Click Print > Instrument Configuration.
13 Verify that the SmartCard 4 firmware revision is 5.30 or higher.