Technical data

2 TOF Software and TOF Firmware Upgrade for Multimode
34 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
25 After reloading software, configure the logical instrument. The TOF and
optionally the LC and ADC are configured into one logical instrument. This
must be done before using the TOF Software system.
a Click Start > Programs > Agilent > TOF Software > Tools > Instrument
Configuration.
b Click Add.
c In the Add Devices window, click OK to accept the default connection
value for the LC/MSD TOF.
d Continue to add instruments until all desired instruments have been
added.
e Change the Instrument Name if desired. This is only used for reporting
purposes.
f Click OK to create the instrument configuration.
g When the instrument is successfully configured, click OK to close the
information box that appears.
26 Check that the HOSTS file were correctly changed. The TOF Software adds
an entry for the TOF and the HPLC modules.
a Use Notepad to open the file HOSTS in the C:\WINNT\System32\drivers\
etc\ folder.
b Under the entry for the local host, check that the entries appear for the
TOF system and LC (if configured).
c Click File > Save to close the file.