Technical data
TOF Software and TOF Firmware Upgrade for Multimode 2
Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide 33
17 Click Next after specifying the destination location for the installed
software. It is recommended to use the default location.
18 Mark the Install BootP check box. SmartCard 4 uses a fixed IP address
(192.168.254.12). Therefore the BootP service is not needed for the TOF
instrument. However, the LAN card for the LC may be configured to use
BootP.
You can also configure the LC with a fixed IP addresses. Please refer to the
CD labeled “Agilent Technologies LAN Card G1369A” (p/n G1369-90000)
for more information.
19 Click Next after specifying the location for data like methods and reports to
be stored. You may wish to put the data on a different drive to make
backups easier. For bundled systems, the default location for TOF data is
D:\TOF_Data. You need to change the value in the location text box to this
folder name.
20 Click Next on the Start Copying Files dialog box that allows you to review
your setup choices.
21 When you are asked to update the list of masses, click Yes. The upgrade
from A.00.00, A.01.00, or A.01.01 to A.02.00 requires the list of masses to be
updated.
22 Click Yes when you are prompted to restart the computer.
23 Connect the House LAN to the computer.
24 Add the MAC address for the LC LAN card to the BootP Manager and assign
the IP address 192.168.254.11 to the LC.
You can also configure the LC with a fixed IP addresses. Please refer to the
CD labeled “Agilent Technologies LAN Card G1369A” (p/n G1369-90000)
for more information.