Technical data

2 TOF Software and TOF Firmware Upgrade for Multimode
32 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
Step 1. Upgrade the TOF Software
1 Back up the following folders in the C:\Program Files\Agilent\TOF Software
folder to a safe location.
DA methods
methods
worklists
Methods and mass lists will be overwritten when the new software is
installed. Also, if you have modified any Agilent supplied files, these files
may be overwritten when the software is upgraded. Back up any modified
Agilent files to a safe location. It is recommended that VBA projects
provided by Agilent should not be modified.
2 Disconnect the House LAN from the computer.
3 Use Windows Add-Remove Programs to remove the previous installation of
TOF Software.
4 When you are asked whether to remove all files and folders, click No. Your
methods, worklists, and data analysis scripts will be preserved.
5 Use Windows Add-Remove Programs to remove the previous installation of
Analyst QS.
6 Use File Explorer and open the CD-ROM drive for the Installation CD. Open
the directory Analyst. Double-click setup.exe.
7 Click Next on the Welcome screen.
8 Click Yes to accept the software license agreement.
9 Click Next after specifying the destination location for the installed
software. It is recommended to use the default location.
10 When the dialog box appears for selecting the data file storage location,
select the location D:\PE Sciex Data\.
11 Click Next to proceed with the rest of the installation.
12 When the dialog box appears to install the o-MALDI server, click Cancel.
This software is not required for the LC/MSD TOF.
13 Click Yes when prompted to reboot the computer.
14 Use File Explorer and open the CD-ROM drive for the Installation CD. Open
the directory G3300. Double-click setup.exe.
15 Click Next on the Welcome screen.
16 16.Click Yes to accept the software license agreement.