Step 7. Install the nebulizer, corona needle, and gas
connections
1 Install the nebulizer on the multimode source spray chamber. Never use a
nebulizer spacer on a multimode source.
2 Connect the 1/8-inch nebulizing gas tubing from the LC/MS instrument
mainframe to the nebulizer gas fitting
3 Connect the LC/MS sample tubing to the nebulizer zero dead volume fitting.
See Figure 20.
4 Connect the LC to the Inlet connection on the TOF instrument. Never
connect LC tubing directly to the Nebulizer. See Figure 20.
5 Install the corona needle. Make sure the corona needle is pushed all the way
in.
Figure 20 Nebulizer, nitrogen gas and corona needle installed
Sample tubing
Nebulizer gas fitting
Corona needle
Inlet