Technical data

1 Installation
22 Agilent Multimode for 6210/6220 TOF Installation and Maintenance Guide
3 Screw the End Cap Assembly into the holder for the end cap. See Figure 18.
Note that the field shaping electrodes are in the 9 o'clock and 6 o'clock
positions.
4 Loosen the end plate screws on each side to adjust the field shaping
electrodes position.
Figure 18 Multimode End Cap Assembly (p/n G1978-60060)
5 Connect the vaporizer heater, APCI high voltage, and HV and control PCA
cables. The APCI heater connector, APCI high voltage connector, and HV
and control PCA connector located on the left side of the instrument
chassis. See Figure 19.
Field shaping electrode,
9 o’clock position
Field shaping electrode,
6 o’clock position